Haefely FP-SURGE 100M2, 3-phase Coupling / Decoupling Network (CDN
Images
  • Haefely FP-SURGE 100M2, 3-phase Coupling / Decoupling Network (CDN

Haefely FP-SURGE 100M2, 3-phase Coupling / Decoupling Network (CDN

Art. nr:

Ask for quotation

 
 

Product description

IEC/EN 61000-4-5, IEC/EN 61000-4-12,
ANSI C62.41 & 62.45
  • Surge Combination Wave, 1.2/50 µs - 8/20 µs
  • Ring Wave, 100 kHz
  • 8 kV impulse voltage
  • Line voltage 690 V AC (phase – phase)
  • 100 A EUT current per phase
The FP-SURGE 100M2 coupling/decoupling network with manual coupling path switching is fully compliant with the standard requirements of IEC/EN 61000-4-5, IEC/EN 61000-4-12, and tests to meet the surge requirements of ANSI/IEEE C62.41 and C62.45. It is specifically designed to meet and exceed the requirements of IEC, EN, and ANSI tests for power line applications.
 
ANSI standards C62.41 and C62.45 contain essentially the same information as the IEC standards, but are based on U.S. experience with AC power lines. No requirements are included for voltage drop across decoupling reactors. 
 
With a selector switch, the FP-SURGE 100M2 can handle different current ranges (0 - 25 A, 25 - 60 A and 60 - 100 A).
 
The FP-SURGE 100M2 EUT output has specially designed high voltage terminals that provide increased personal protection with respect to the high voltage pulse.