The TBL5016-2 is a Line Impedance Stabilization Network for the measurement of line-conducted interference within the range of 9kHz to 30MHz, according to the CISPR 16-1-2 standard.
The device is designed for testing single phase, AC-powered equipment with supply voltages up to maximum 240V and 16A.
Conducted noise can be measured on the phase and on the neutral conductor. The TBL5016-2 is equipped with a switchable limiter/attenuator and an artificial hand connection.
The device is available with country-specific DUT connectors.
Specifications:
Frequency range: 9 kHz to 30 MHzImpedance: 50 Ω ║ (50 µH + 5 Ω)Artificial hand: 220 pF + 511 ΩSwitchable PE: 50 Ω ║ 50 µHLimiter / attenuator: 150 kHz to 30 MHz; 10 dBAir core inductorsLine voltage: max. 240V / 50 – 60 Hz, CAT ‖‖Max. current: 16A @ 23°CDUT socket: country specificMeasurement connector: 50 Ω BNCPower connector: IEC 60320 C19Operating Temperature Range: +5°C … + 40°C; 5% to 80% RHSafety: Safety Class I, IEC 1010-01